DERMATOLOGY 1. First focus normally on a slide with one eye
HOW TO SET UP YOUR MICROSCOPE FOR DERMATOLOGIC SAMPLES
looking through the  xed (non-adjustable) eyepiece.
N. Tim1 2. Next, close your  rst eye and then correct
1University of Edinburgh, Royal Dick School of Veterinary Studies, Roslin, United Kingdom
HOW TO SET UP YOUR MICROSCOPE FOR DERMATOLOGIC SAMPLES
Dr Tim Nuttall
Royal (Dick) School of Veterinary Studies, University of Edinburgh, Easter Bush Campus, Roslin, UK.
tim.nuttall@ed.ac.uk
It is very important to set up your microscope correctly. Incorrect set up results in poor-quality images, eye strain, headaches and motion sickness. This discourages use and compromises the diagnosis.
Correct microscope setup
1. Alter the separation between the eyepieces until you see a single, circular image. You can make a note of the distance for quick reference in the future.
2. Most microscopes have wide- eld eyepiece lenses suitable for people that wear glasses. If you do, fold the rubber eyecups down and rest your glasses against them. If you don’t, fold the eyecups out and rest your orbit against them. Your eyes will now be the correct distance from the lenses.
3. Focus to get a sharp image with relaxed eyes; i.e. focus on in nity and don’t strain to get the image in focus.
4. Adjust the focus between the eyepieces for your eyesight using the adjustable (independently focussing) eyepiece. The adjustment may be on the eyepiece or on the housing (i.e. on the microscope body itself). There will be a 0 or other mark – when this is aligned both eyepieces are confocal, which is appropriate for perfect or perfectly corrected vision. Most people have slight differences between their eyes.
the focus for your other eye, if necessary, by adjusting the other eyepiece (using its independent focus).
3. You will now have the two binocular images in perfect focus for each eye. This avoids eyestrain and blurring of edges.
5. Koehler illumination focuses the light source on the slide, giving the optical illusion that the light comes from the sample. This avoids transillumination and gives you the best light balance and image quality. It is very easy to set up:
1. Focus on a slide.
2. Close the light source lens diaphragm so that you can see the edges (they look like a hexagon or octagon) in the image  eld using the x4 or x10 objective lenses.
3. Focus the condenser until edges of the diaphragm are sharp.
4. If necessary, adjust the condenser centring screws until the light source is centred in the  eld of view.
5. Open the diaphragm so that it is no longer visible and there is even illumination of the  eld of view.
6. Some older and/or cheaper microscopes do not have a light source lens diaphragm – hold a thin piece of card or paper, a paper-clip, or the tip of a pen or pencil against the light source and focus the condenser until you see a sharp image outline in the image  eld.
6. You can now adjust the iris diaphragm to give the clearest image for each lens:
1. a. For parasites and dermatophytes close the iris diaphragm – the image is poorer but the increased contrast makes the parasites or fungi stand out better on unstained samples.
b. For cytology of stained preparations under high power, open the diaphragm to reduce contrast and improve the detail of the nucleus, cytoplasm, granules and microorganisms.